June 2024 Microscope Status Update
A benchmarking was performed at the magnification of 130kx on Jun 25 for the Titan Krios.
1060 exposures with a 5A CTF filter and 482,111 particles were used for processing. A resolution of 1.90A was obtained. The pixel size is around 0.93A.
The Glacios benchmarking was conducted at a magnification of 120kx on Jun 28. The beam size problem has been fixed. 265 exposures with a 4A CTF filter and 202,200 particles were used for processing. A resolution of 2.43A was obtained. The pixel size is around 1.19A.
The Aquilos is operational with a fixed chilling water problem and ion beam instability.