July 2024 Microscope Status Update

A benchmarking was performed at the magnification of 130kx on July 17 for the Titan Krios. 1025 exposures with a 5A CTF filter and 442,621 particles were used for processing. A resolution of 1.90A was obtained. The pixel size is around 0.93A. The Glacios benchmarking was conducted at a magnification of 120kx on July 24. 739 exposures with a 5.5A…

Read more

May 2024 Microscope Status Update:

A benchmarking was performed at magnification 130kx on May 30 for the Titan Krios. 614 exposures with a 5A CTF filter and 160,880 particles were used for processing. A resolution of 1.94A was obtained. The pixel size is around 0.93A. The Glacios benchmarking was conducted at a magnification of 120kx on May 20. 753 exposures with a 5A CTF filter…

Read more

April 2024 Microscope Status Update:

A benchmarking was performed at a magnification 215kx on April 15 for the Titan Krios. 2293 exposures with a 6A CTF filter and 329002/346312 particles were used for processing. A resolution of 1.73A was obtained. The pixel size is around 0.56A. The Glacios benchmarking was conducted with an Au 0.6 grid at a magnification of 120kx on April 15 after…

Read more

March 2024 Microscope Status Update

We performed a benchmarking on March 29 for the Titan Krios. At magnification 130kx, we collected 755 exposures and used 711 with a 5A CTF filter. We used 329002/346312 particles and obtained a resolution of 1.89A. The pixel size is around 0.93A. The Glacios benchmarking was done on March 27. At magnification 120kx, we collected 814 exposures and used 786…

Read more